• 282 Downloads • Abstract This study assessed the validity of the conversion from percentage depth dose (PDD) to tissue maximum ratio (TMR) using BJR Supplement 25 data for flattened and flattening filter free (FFF) beams. Vst Plugins For Cubase 5. PDD and TMR scans for a variety of field sizes were measured in water using a Sun Nuclear Corporation 3D SCANNER™ on a Varian TrueBeam linear accelerator in 6 MV, 10 MV and 6 MV FFF beams. Download Revista Satiricon Rar. The BJR Supplement 25 data was used to convert the measured PDDs to TMRs and these were compared with the directly measured TMR data. The TMR plots calculated from PDD were within 1% for the 10 MV and 6 MV flattened beams, for field sizes 3 cm × 3 cm to 40 cm × 40 cm inclusive, at depths measured beyond the depth of maximum dose.

Bjr Supplement 25 Pdf

DOC-Live: british journal of radiology supplement 25 (bjr 25 1996) - Online Free Unlimited pdf document search and download.

The disagreement between the measured and calculated TMR plots for the 6 MV FFF beam increased with depth and field size to a maximum of 1.7% for a 40 cm × 40 cm field. The results found in this study indicate that the BJR Supplement 25 data should not be used for field sizes larger than 20 cm × 20 cm at depths greater than 15 cm for the 6 MV FFF beam. It is advised that PDD to TMR conversion for FFF beams should be done with phantom scatter ratios appropriate to FFF beams, or the TMR should be directly measured if required.